Direct potentiometry(Nova-l Analyzer)와 Flame photometry(FLM3)에 의한 Sodium과 potassium의 측정치 비교
김서혜 ( Shu Hae Kim ) 대한임상검사과학회 1984 대한임상검사과학회지(KJCLS) Vol.16 No.1
The relationship between direct-potentiometric (Nova-1) and f1ame photometric (FLM3) measurements of Na & K was evaluated in samples of human sera and urine. The results were summarized as follows: 1. When measuring Na & K, two instruments gave linear responses and good precision. But especially in the blood mode, the precision of Nova-1 was better than FLM3. 2. With the instuments in the blood mode the correlation coefficient for Na was 0.869 and for K , 0.977. In the urine mode, correlation coefficient for Na was 0.994 and for K, 0.990. 3. With the lipemic serum (triglyceride:> 200mg/dl) the correlation coefficient of Na & K in the two insrtuments was 0.749 and 0.878, respectively.