김형균,성동근,조경미,김상범,김재영,최준호,Kim, Hyong-Gyun,Sung, Dong-Keon,Cho, Kyong-Mi,Kim, Sang-Beom,Kim, Jae-Young,Choi, Jun-Ho 대한디지털의료영상학회 2009 대한디지털의료영상학회논문지 Vol.11 No.2
This study, "The study about performance evaluations of mobile cover for X-ray's diffusion and distribution in mobile radiation" is based on the rules of mobile defense apparatus for radiation producer in 2006. To use the mobile cover for X-ray for diagnosis has been compulsory in common wards except operation rooms, emergency rooms and intensive care units. we have confirmed the effect in arbitrary shielding material after Qualitiy Control was carried out for accuracy in an experiment of mobile photographing equipment. The performance evaluation was conducted with the fabrics of selenium, 0.2 mmPb, 0.1 mmPb and aluminiums. Considering the result, we choosed 0.1 mmPb and attached cover to mobile photographing equipment. We have finished making the cover after drew up the draft to attach cover to mobile photographing equipment through the modeling and the structural analysis. the process of the study is that we assembled the manufactured structures and carried out the practical experiment to take the photograph after attaching the fabric of 0.1 mmPb to mobile photographing equipment. It is need of additional thesises hereafter that we compare the result between the part to improve for safety besides convenience in photographic experiment about clinical radiation and the effect of covering the diffusion in condition attached the cover.