교대부양 설정압력 변화가 인체조직의 동적인 관류특성에 미치는 영향
원병희(Byeong Hee Won),송창섭(Changseop Song) 대한인간공학회 2010 大韓人間工學會誌 Vol.29 No.6
The quantitative effectiveness of powered support surfaces such as APAM in preventing and treating pressure ulcers has not been sufficiently evaluated because of uncertainty of alternating pressure load input and lack of interpretation of dynamic perfusion characteristics of soft tissue. The aim was to verify the dynamic loading effects to sacral tissue perfusion characteristics from alternating set pressure changes. We developed integrated experiment system to supply alternating load to supinely positioned sacrum and concurrently measured TcPO₂, TcPCO₂ and air cell pressure. Ten aged subjects (5 female, 5 male) were tested with alternating set pressure 20, 30, 40,50 and 60㎜Hg. From the dynamic perfusion response eight characteristic parameters were proposed such as average, minimum, maximum and perfusion range regarding to TcPO₂ and TcPCO₂. A one-way ANOVA was carried out to determine whether the manipulation of alternating set pressure had any effect on TcPO₂ and TcPCO₂. From the dynamic tissue perfusion response we found mean TcPO₂ decreased exponentially as alternating pressure load increased and perfusion range varied mainly because of minimum level change of TcPO₂. And perfusion range of TcPCO₂ affected by increase of maximum value of TcPCO₂. From the results we can get more strict insights about actual physiological dynamic tissue perfusion mechanism under alternating pressure load.