Auto-Analyzer Ⅱ法과 Dextrostix法에 依한 血糖定量에 關하여
丁英鎭,蔡範錫,南瀧江 충남대학교 자연과학연구소 1976 學術硏究誌 Vol.3 No.2
Recently the utility of the Dextrostix blood sugar estimation method has been greatly jncreased by the use of the reflectance meter which was devided for quantitative reading of color change in Dextrostix. A comparison study of 300 blood sugar determinations were done using blood serum from in patient and patients attanding diabetic clinic. The Dextro-stix-Reflectance meter method was used and the results were compared to an Auto-Analyzer Ⅱ method. The correlation of these blood sugar results with those obtained with an Auto-Analyzer Ⅱ method was very high (r=0. 996) and the replication of the Dextrostix-Reflectance meter method by the different experimenters was also satisfactory. These results indicate that Dextrostix-Reflectance meter is a simple 2nd reliable for estimating the blood sugar level if procedure are accurately performed.