http://chineseinput.net/에서 pinyin(병음)방식으로 중국어를 변환할 수 있습니다.
변환된 중국어를 복사하여 사용하시면 됩니다.
혈액배양에서 VITEK MS와 VITEK 2 System을 이용한 신속 항생제 감수성 시험의 유용성 평가
박강균,유영빈,육근돌,김상하,김성현,김영권,Park, Kang-Gyun,Yu, Young-Bin,Yook, Keundol,Kim, Sang-Ha,Kim, Sunghyun,Kim, Young Kwon 대한임상검사과학회 2017 대한임상검사과학회지(KJCLS) Vol.49 No.3
The results of rapid antimicrobial susceptibility test (AST) in blood cultures were obtained by inoculating the bacteria directly into the VITEK MS and the VITEK 2 systems without subculturing in the blood culture positive medium. The obtained results were compared with the results using a standard method to evaluate their reliability and accuracy. The direct AST results in blood culture positive specimens were 97.9% (1,936/1,978), consistent with the standard AST results. Gram-positive bacteria showed a concordance rate of 97.2% (1,051/1,081), a very major error rate of 0.5% (5/1,081), a major error rate of 0.1% (1/1,081), and a minor error rate of 2.2% (24/1,081). Staphylococcus epidermidis was the main cause of discordance, and gentamicin (N=9) and fusidic acid (N=8) showed high errors. The overall concordance rate and minor error among the Gram-negative bacteria were 98.6% (885/897) and 1.4% (12/897), respectively. Escherichia coli and Pseudomonas aeruginosa were the major causative bacteria of Gram-negative bacteria. Among them, amoxicillin/clavulanic acid (N=3) showed high error. Direct AST met the CLSI criteria and shortened the reporting time by 24 hours; however, we found that there was a need to perform an addition test via disk diffusion for antimicrobials with very large errors. These results suggest that the method of direct AST in blood culture positive medium may be very useful in efficiently treating patients.