Bacchi, Atais,Regalin, Alexandre,Bhering, Claudia Lopes Brilhante,Alessandretti, Rodrigo,Spazzin, Aloisio Oro The Korean Academy of Prosthodonitics 2015 The Journal of Advanced Prosthodontics Vol.7 No.5
PURPOSE. The purpose of this study was to evaluate the influence of tightening technique and the screw coating on the loosening torque of screws used for Universal Abutment fixation after cyclic loading. MATERIALS AND METHODS. Forty implants (Titamax Ti Cortical, HE, Neodent) (n=10) were submerged in acrylic resin and four tightening techniques for Universal Abutment fixation were evaluated: A - torque with 32 Ncm (control); B - torque with 32 Ncm holding the torque meter for 20 seconds; C - torque with 32 Ncm and retorque after 10 minutes; D - torque (32 Ncm) holding the torque meter for 20 seconds and retorque after 10 minutes as initially. Samples were divided into subgroups according to the screw used: conventional titanium screw or diamond like carbon-coated (DLC) screw. Metallic crowns were fabricated for each abutment. Samples were submitted to cyclic loading at $10^6$ cycles and 130 N of force. Data were analyzed by two-way ANOVA and Tukey's test (5%). RESULTS. The tightening technique did not show significant influence on the loosening torque of screws (P=.509). Conventional titanium screws showed significant higher loosening torque values than DLC (P=.000). CONCLUSION. The use of conventional titanium screw is more important than the tightening techniques employed in this study to provide long-term stability to Universal Abutment screws.
Atais Bacchi,Alexandre Regalin,Claudia Lopes Brilhante Bhering,Rodrigo Alessandretti,Aloisio Oro Spazzin 대한치과보철학회 2015 The Journal of Advanced Prosthodontics Vol.7 No.5
PURPOSE. The purpose of this study was to evaluate the influence of tightening technique and the screw coating on the loosening torque of screws used for Universal Abutment fixation after cyclic loading. MATERIALS AND METHODS. Forty implants (Titamax Ti Cortical, HE, Neodent) (n=10) were submerged in acrylic resin and four tightening techniques for Universal Abutment fixation were evaluated: A ? torque with 32 Ncm (control); B ? torque with 32 Ncm holding the torque meter for 20 seconds; C ? torque with 32 Ncm and retorque after 10 minutes; D ? torque (32 Ncm) holding the torque meter for 20 seconds and retorque after 10 minutes as initially. Samples were divided into subgroups according to the screw used: conventional titanium screw or diamond like carbon-coated (DLC) screw. Metallic crowns were fabricated for each abutment. Samples were submitted to cyclic loading at 106 cycles and 130 N of force. Data were analyzed by two-way ANOVA and Tukey’s test (5%). RESULTS. The tightening technique did not show significant influence on the loosening torque of screws (P=.509). Conventional titanium screws showed significant higher loosening torque values than DLC (P=.000). CONCLUSION. The use of conventional titanium screw is more important than the tightening techniques employed in this study to provide long-term stability to Universal Abutment screws.