CT performance phantom을 이용한 cone beam형 전산화단층영상의 특성
한충완,김규태,최용석,황의환 대한구강악안면방사선학회 2007 Imaging Science in Dentistry Vol.37 No.3
Purpose : To evaluate the characteristics of (widely used) cone beam computed tomography (CBCT) images. Materials and Methods : Images were obtained with CT performance phantoms (The American Association of Physicists in Medicine; AAPM). CT phantom as the destination by using PSR 9000N™ dental CT system (Asahi Roentgen Ind. Co., Ltd., Japan) and i-CAT CBCT (Imaging Science International Inc., USA) that have different kinds of detectors and field of view, and compared these images with the CT number for linear attenuation, contrast resolution, and spatial resolution. Results : CT number of both PSR 9000N™ dental CT system and i-CAT CBCT did not conform to the base value of CT performance phantom. The contrast of i-CAT CBCT is higher than that of PSR 9000N™ dental CT system. Both contrasts were increased according to thickness of cross section. Spatial resolution and shapes of reappearance was possible up to 0.6 mm in PSR 9000N™ dental CT system and up to 1.0 mm in i-CAT CBCT. Low contrast resolution in region of low contrast sensitivity revealed low level at PSR 9000N™ dental CT system and i-CAT CBCT. Conclusion : CBCT images revealed higher spatial resolution, however, contrast resolution in region of low contrast sensitivity was the inferiority of image characteristics.