http://chineseinput.net/에서 pinyin(병음)방식으로 중국어를 변환할 수 있습니다.
변환된 중국어를 복사하여 사용하시면 됩니다.
의료용 약물주입펌프의 조작스위치 내구성 검증 장비 개발 및 신뢰성입증시험
오상민(Sangmin Oh),신정훈(Junghun Shin),정봉수(Bongsu Jung),정재훈(Jaehoon Jung),김태형(Taehyung Kim),박근묵(Geunmook Park),서수원(Soowon Seo) 대한기계학회 2015 대한기계학회 춘추학술대회 Vol.2015 No.11
Through market analysis it has been reported that many issues have occurred concerning the reliability of the medical drug infusion pump’s performance in administering drugs into the human body. A test method that demonstrates the reliability of the drug infusion pump is critical insofar as it is directly connected to the patient’s life, and so in this study we have analyzed its failure trends of each part through the analysis of the failure mode of the drug infusion pump, and we have developed equipment as well as a test method that allows demonstration of the reliability of the control switch, with which frequent failures have been observed as being expected to occur. On this basis, a lifespan test has been conducted on the switch, and a study has been conducted in order to demonstrate the reliability of the drug infusion pump by establishing a guaranteed lifetime.