http://chineseinput.net/에서 pinyin(병음)방식으로 중국어를 변환할 수 있습니다.
변환된 중국어를 복사하여 사용하시면 됩니다.
A novel reference model for dental scanning system evaluation: analysis of five intraoral scanners
Irina Karakas-Stupar,Nicola Ursula Zitzmann,Tim Joda 대한치과보철학회 2022 The Journal of Advanced Prosthodontics Vol.14 No.2
PURPOSE. The aim of this in vitro study was to investigate the accuracy (trueness and precision) of five intraoral scanners (IOS) using a novel reference model for standardized performance evaluation. MATERIALS AND METHODS. Five IOSs (Medit i500, Omnicam, Primescan, Trios 3, Trios 4) were used to digitize the reference model, which represented a simplified full-arch situation with four abutment teeth. Each IOS was used five times by an experienced operator, resulting in 25 STL (Standard Tessellation Language) files. STL data were imported into 3D software (Final Surface®) and examined for inter- and intra-group analyses. Deviations in the parameter matching error were calculated. ANOVA F-test and Kruskal-Wallis test were applied for inter-group comparisons (α = .05); and the coefficient of variation (CV) was calculated for intra-group comparisons (in % ± SD). RESULTS. Primescan (matching error value: 0.015), Trios 3 (0.016), and Trios 4 (0.018) revealed comparable results with significantly higher accuracy compared to Medit i500 (0.035) and Omnicam (0.028) (P < .001). For intra-group comparison, Trios 4 demonstrated the most homogenous results (CV 15.8%). CONCLUSION. The novel reference model investigated in this study can be used to assess the performance of dental scanning technologies in the daily routine setting and in research settings.