http://chineseinput.net/에서 pinyin(병음)방식으로 중국어를 변환할 수 있습니다.
변환된 중국어를 복사하여 사용하시면 됩니다.
Mustafa Soylak,Sevtap Yigit 한국공업화학회 2015 Journal of Industrial and Engineering Chemistry Vol.24 No.-
A separation–preconcentration procedure for germanium has been presented. The optimum conditions including pH, sample volume, sample-eluent flow rates, sample and matrix effects for adsorption of germanium(IV) on polysulfone membrane filter were investigated prior to spectrophotometric determination of germanium as Ge(IV)-9-phenyl-3-fluorone complex. Ge(IV) was adsorbed on membrane at pH 4.0 and was eluted from membrane with phosphoric acid. The accuracy was checked by determination of germanium content of GBW 07402 soil certified reference material. The detection and quantification limits were found as 2.1 and 7.0 ng/L, respectively. The method was applied for determination of germanium in soil and water samples.