Novel Thallium(I)-Selective Membrane Electrode Based on a Podal Ligand
Ganjali, Mohammad Reza,Pourjavid, Mohammad Reza,Mouradzadegun, Arash,Hosseini, Morteza,Mizani, Farhang Korean Chemical Society 2003 Bulletin of the Korean Chemical Society Vol.24 No.11
A PVC-based membrane electrode for thallium(I) ions based on 1,21,23,25-tetramethyl-2,20: 3,19-dimetheno-[H, 2] H, 23H, 25H-bis-[1,3] dioxocino[5,4-i:5',4'-i] benzo [1,2-d: 5.4-d'] bis [1,3] benzodioxocin(II) has been prepared. The electrode displays a linear dynamic range of $1.0{\times}10^{-1}-1.0{\times}10^{-5}$ M, with a Nernstian slope of $59.8{\pm}0.2\;mV\;{decad^-1}$, and a detection limit $5.0{\times}10^{-6}$ M. It has a very fast response time of<10 s and can be used for at least ten weeks without a considerable divergence in potentials. This electrode revealed comparatively good selectivity with respect to alkali, alkaline earth, and some transition and heavy metal ions and was effective in a pH range of 2.0-10.0. It was used as an indicator electrode in potentiometric titration of thallium ion with sulfide ion.