http://chineseinput.net/에서 pinyin(병음)방식으로 중국어를 변환할 수 있습니다.
변환된 중국어를 복사하여 사용하시면 됩니다.
Taher, Mohammad Ali,Mostafavi, Ali,Afzali, Darush,Rezaeipour, Ebrahim Korean Chemical Society 2004 Bulletin of the Korean Chemical Society Vol.25 No.8
This work assesses the potential of natural Analcime Zeolite as an adsorbent for preconcentration of lead (II) traces. Lead is quantitatively retained on 2-(5-bromo-2-pyridylazo)-5-diethylaminophenol by column method with Analcime in the pH range of 5-6.5 and 2 mL $min^{?1}$ flow rate. Lead was removed from the column with 10.0 mL of 4 M hydrochloric acid and was determined by anodic stripping differential pulse voltammetry. 0.5ppb detection limit was obtained and linear dynamic range was 3 to $1.2{\times}10^5$ ppb in final solution with correlation coefficient of 0.999 and relative standard deviation of ${\pm}$ 1.2% (for eight replicate determination of 2.5 ${\mu}g\;mL^{?1}$ of lead). Various parameters such as the effect of pH, flow rate, instrumental conditions and interferences of some ions on the determination of lead have been studied in detail for optimization of conditions. The method was successfully applied for determination of lead in various samples.