원병희(B. H. Won),전경진(K. J. Chun) 한국정밀공학회 2006 한국정밀공학회 학술발표대회 논문집 Vol.2006 No.5월
A patient table considering x-ray transparency, mechanical safety and compact multi-axis moving mechanism has been developed. The goal of medical imaging technology is to keep radiation exposure of patients during x-raying to a minimum. In order to obtain clear pictures at low dose, however, the x-ray table which supports the patient must be sufficiently permeable to radiation to allow good image resolution. The table top is made of low density foam for x-ray transparent effective area and structural aluminum plate to connect moving mechanism under the table, covered with thin carbon fiber. This sandwich construction is very rigid and lightweight, so the table top can handle relatively heavy load comparing to its cantilevered structure which is unavoidable as long as cooperate with C-arm radiography. To verify the design results finite element static analysis and experimental tests have been done. According to the verification the results well satisfy certification guide lines as a medical device.