Cassette, Philippe,Altzitzoglou, Timotheos,Antohe, Andrei,Rossi, Mario,Arinc, Arzu,Capogni, Marco,Galea, Raphael,Gudelis, Arunas,Kossert, Karsten,Lee, K.B.,Liang, Juncheng,Nedjadi, Youcef,Oropesa Verd Elsevier 2018 Applied radiation and isotopes Vol.134 No.-
<P><B>Abstract</B></P> <P>A comparison of calculations of the activity of a <SUP>3</SUP>H<SUB>2</SUB>O liquid scintillation source using the same experimental data set collected at the LNE-LNHB with a triple-to-double coincidence ratio (TDCR) counter was completed. A total of 17 laboratories calculated the activity and standard uncertainty of the LS source using the files with experimental data provided by the LNE-LNHB. The results as well as relevant information on the computation techniques are presented and analysed in this paper. All results are compatible, even if there is a significant dispersion between the reported uncertainties. An output of this comparison is the estimation of the dispersion of TDCR measurement results when measurement conditions are well defined.</P> <P><B>Highlights</B></P> <P> <UL> <LI> Comparison of calculations of the activity of a <SUP>3</SUP>H<SUB>2</SUB>O LS source from the same experimental data set. </LI> <LI> Data set from a triple-to-double coincidence ratio (TDCR) counter. </LI> <LI> 17 laboratories participated in this exercise. </LI> <LI> Results and relevant information on the computation techniques are presented and analysed. </LI> <LI> All results are compatible but significant dispersion of the reported uncertainties. </LI> </UL> </P>