Quality assurance for fine particle requires conditioning of filters at 20∼23℃ and 30∼40% relative humidity for 24hr prior to gravimetric analysis. The reason for these requirements is that the accuracy or reliability of filter weighting is asso...
http://chineseinput.net/에서 pinyin(병음)방식으로 중국어를 변환할 수 있습니다.
변환된 중국어를 복사하여 사용하시면 됩니다.
https://www.riss.kr/link?id=A60123800
2011
Korean
405
학술저널
117-122(6쪽)
0
상세조회0
다운로드다국어 초록 (Multilingual Abstract)
Quality assurance for fine particle requires conditioning of filters at 20∼23℃ and 30∼40% relative humidity for 24hr prior to gravimetric analysis. The reason for these requirements is that the accuracy or reliability of filter weighting is asso...
Quality assurance for fine particle requires conditioning of filters at 20∼23℃ and 30∼40% relative humidity for 24hr prior to gravimetric analysis. The reason for these requirements is that the accuracy or reliability of filter weighting is associated with and can be affected by several factors including temperature, humidity, light, draft, and electrostatic charge. In this study, concentration of respirable suspended particulate by gravimetric method was measured in smoking room by three kinds of filters (PVC, MCE and PTFE) to quantify the effect of electrostatic charge. Among the filters, MCE filter was statistically affected by electrostatic charge (p= 0.002).
디지털 음악 및 아날로그 음악이 악력에 미치는 영향 평가