      Possibility of Interchanging Patients for Beam-Matched Linear Accelerators from the Same Vendor

      한글로보기

      https://www.riss.kr/link?id=A106406379

      • 0

        상세조회
      • 0

        다운로드
      서지정보 열기
      • 내보내기
      • 내책장담기
      • 공유하기
      • 오류접수

      부가정보

      다국어 초록 (Multilingual Abstract)

      The purpose of this study is to investigate the possibility of interchanging patients on beam-matched linear accelerators (linacs), by evaluating the dosimetric variation and the dose delivery accuracy of volumetric modulated arc therapy (VMAT). Two V...

      The purpose of this study is to investigate the possibility of interchanging patients on beam-matched linear accelerators (linacs), by evaluating the dosimetric variation and the dose delivery accuracy of volumetric modulated arc therapy (VMAT). Two VitalBeam linacs installed recently in our center were included in this study. For dosimetric data comparison, beam data, such as percentage depth doses (PDDs), beam profiles, output factors, multi-leaf collimator (MLC) leaf transmission factors, and the dosimetric leaf gap (DLG), for 6-MV and 10-MV beams were measured on both linacs. Sixty patients (20 head and neck, 20 lung, and 20 prostate cancers) were selected to evaluate the dosimetric parameters and the delivery accuracy of the VMAT plans created for various treatment sites. All the VMAT plans were generated using the Eclipse treatment planning system (TPS) with the same optimizer algorithm. The dose differences between two beam-matched linacs for planning target volumes (PTVs) and organs at risk (OARs) were recorded and analyzed. Patient-specific dosimetric quality assurance (DQA) using the Electronic Portal Image Device (EPID) were performed, and the measured planar doses were compared with those calculated by using the TPS though absolute dose gamma evaluation using 2%/2 mm and 2%/3 mm criteria. The differences in all dosimetric beam data between the two beam-matched linacs were all within 1% for the 6-MV and the 10-MV beams. For all 60 cases of three treatment sites, no notable differences were observed in the dosimetric comparison and the DVH variation between the beam-matched linacs. The average differences in dosimetric parameters of PTV and OARs were all within 1%. Furthermore, all passing rates of measured doses on the beam-matched linacs were higher than 93% and 97% for the 2%/2 mm and the 2%/3 mm criteria, respectively. Our overall dosimetric variation in beam data and VMAT plans remains within the tolerance of clinical acceptability. Pretreatment verification results were also clinically acceptable, showing good agreement with the calculated doses. Therefore, this study demonstrated the possibility of swapping patients in a VMAT plan without any modification and re-planning between two beam-matched linacs.

      더보기

      참고문헌 (Reference)

      1 R. Nath, 21 : 1094-, 1994

      2 J. Hrbacek, 34 : 2917-, 2007

      3 D. Sj¨ostr¨om, 48 : 192-, 2009

      4 G. P. Beyer, 14 : 273-, 2013

      5 H. A. Gay, 83 : e353-, 2012

      6 Z. Xu, 20 : 68-, 2018

      7 J. R. Bhangle, 36 : 176-, 2011

      8 R. J. Watts, 26 : 2581-, 1999

      9 E. M. Attalla, 13 : 89-, 2014

      10 C. Krishnappan, 11 : 423-, 2018

      1 R. Nath, 21 : 1094-, 1994

      2 J. Hrbacek, 34 : 2917-, 2007

      3 D. Sj¨ostr¨om, 48 : 192-, 2009

      4 G. P. Beyer, 14 : 273-, 2013

      5 H. A. Gay, 83 : e353-, 2012

      6 Z. Xu, 20 : 68-, 2018

      7 J. R. Bhangle, 36 : 176-, 2011

      8 R. J. Watts, 26 : 2581-, 1999

      9 E. M. Attalla, 13 : 89-, 2014

      10 C. Krishnappan, 11 : 423-, 2018

      11 M. K. Krishna, 8 : e28-, 2012

      12 Y. L. Kim, 64 : 1207-, 2014

      13 백태성, "Feasibility Study of Patient Specific Quality Assurance Using Transit Dosimetry Based on Measurement with an Electronic Portal Imaging Device" 한국의학물리학회 28 (28): 54-60, 2017

      더보기

      분석정보

      View

      상세정보조회

      0

      Usage

      원문다운로드

      0

      대출신청

      0

      복사신청

      0

      EDDS신청

      0

      동일 주제 내 활용도 TOP

      더보기

      주제

      연도별 연구동향

      연도별 활용동향

      연관논문

      연구자 네트워크맵

      공동연구자 (7)

      유사연구자 (20) 활용도상위20명

      인용정보 인용지수 설명보기

      학술지 이력

      학술지 이력
      연월일 이력구분 이력상세 등재구분
      2023 평가예정 해외DB학술지평가 신청대상 (해외등재 학술지 평가)
      2020-01-01 평가 등재학술지 유지 (해외등재 학술지 평가) KCI등재
      2011-01-01 평가 등재학술지 유지 (등재유지) KCI등재
      2009-01-01 평가 등재학술지 유지 (등재유지) KCI등재
      2007-01-01 평가 SCI 등재 (등재유지) KCI등재
      2005-01-01 평가 등재학술지 유지 (등재유지) KCI등재
      2002-07-01 평가 등재학술지 선정 (등재후보2차) KCI등재
      2000-01-01 평가 등재후보학술지 선정 (신규평가) KCI등재후보
      더보기

      학술지 인용정보

      학술지 인용정보
      기준연도 WOS-KCI 통합IF(2년) KCIF(2년) KCIF(3년)
      2016 0.47 0.15 0.31
      KCIF(4년) KCIF(5년) 중심성지수(3년) 즉시성지수
      0.26 0.2 0.26 0.03
      더보기

      이 자료와 함께 이용한 RISS 자료

      나만을 위한 추천자료

      해외이동버튼