      호흡동조방사선치료를 위한 Trigger mode 투시영상 획득 시 호흡 속도에 따른 정확성 평가 - Phantom Study = Evaluation of the Accuracy and usability of Trigger mode in Respiratory Gated Radiation Therapy

      한글로보기

      https://www.riss.kr/link?id=A108109058

      • 0

        상세조회
      • 0

        다운로드
      서지정보 열기
      • 내보내기
      • 내책장담기
      • 공유하기
      • 오류접수

      부가정보

      다국어 초록 (Multilingual Abstract)

      Purpose : The purpose of this study is to evaluate the accuracy and usefulness of the Trigger mode for the Respiratory Gated Radiation Therapy (RGRT) Materials and methods : A QUASAR respiratory phantom that inserted a 3 mm fiducial marker (a gold marker) was used to estimate the accuracy of the Trigger mode. And the 20 bpm was used as reference respiration rate in this study. The marker that placed at the center of the phantom was contoured, and the lower threshold of a gating window was fixed at 2.0 mm using an OBI with Truebeam STx<sup>TM</sup>. The upper threshold was measured every 0.5 mm from 1.0 mm to 3.0 mm. The respiration rates were changed every 10 bpm from 10 bpm to 60 bpm. We repeatedly measured five times to check the error rate of the trigger mode in the same condition. Result : The differences of a distance from a peak phase to upper threshold, 1.0 to 3.0 mm at a 20 bpm as a reference for 3 days in a row were 0.68±0.05 mm, 0.91±0.03 mm, 1.23±0.03 mm, 1.42±0.04 mm, and 1.66±0.06 mm, respectively. Measurement result of changes in respiratory rate compared to baseline respiratory rate in maximum absolute difference. The coefficient of determination (R2) to estimate the correlation between the respiration velocity and variation of absolute difference was on average 0.838, 0.887, 0.770, 0.850, and 0.906. The p-values of all the variables were below 0.05. Conclusion : Using Trigger mode during respiratory gated radiation therapy (RGRT), accuracy and usefulness of trigger mode at reference breathing rate were confirmed. However, inaccuracies depending on the rate of breathing it could be uncertain in case of respiration rate is faster than 20 bpm as a standard respiration rate compared to slower than 20 bpm. Consequently, when conducting a RGRT using the trigger mode, real time monitoring is required with well educated respiration.
      번역하기

      Purpose : The purpose of this study is to evaluate the accuracy and usefulness of the Trigger mode for the Respiratory Gated Radiation Therapy (RGRT) Materials and methods : A QUASAR respiratory phantom that inserted a 3 mm fiducial marker (a gold mar...

      Purpose : The purpose of this study is to evaluate the accuracy and usefulness of the Trigger mode for the Respiratory Gated Radiation Therapy (RGRT) Materials and methods : A QUASAR respiratory phantom that inserted a 3 mm fiducial marker (a gold marker) was used to estimate the accuracy of the Trigger mode. And the 20 bpm was used as reference respiration rate in this study. The marker that placed at the center of the phantom was contoured, and the lower threshold of a gating window was fixed at 2.0 mm using an OBI with Truebeam STx<sup>TM</sup>. The upper threshold was measured every 0.5 mm from 1.0 mm to 3.0 mm. The respiration rates were changed every 10 bpm from 10 bpm to 60 bpm. We repeatedly measured five times to check the error rate of the trigger mode in the same condition. Result : The differences of a distance from a peak phase to upper threshold, 1.0 to 3.0 mm at a 20 bpm as a reference for 3 days in a row were 0.68±0.05 mm, 0.91±0.03 mm, 1.23±0.03 mm, 1.42±0.04 mm, and 1.66±0.06 mm, respectively. Measurement result of changes in respiratory rate compared to baseline respiratory rate in maximum absolute difference. The coefficient of determination (R2) to estimate the correlation between the respiration velocity and variation of absolute difference was on average 0.838, 0.887, 0.770, 0.850, and 0.906. The p-values of all the variables were below 0.05. Conclusion : Using Trigger mode during respiratory gated radiation therapy (RGRT), accuracy and usefulness of trigger mode at reference breathing rate were confirmed. However, inaccuracies depending on the rate of breathing it could be uncertain in case of respiration rate is faster than 20 bpm as a standard respiration rate compared to slower than 20 bpm. Consequently, when conducting a RGRT using the trigger mode, real time monitoring is required with well educated respiration.

      더보기

      분석정보

      View

      상세정보조회

      0

      Usage

      원문다운로드

      0

      대출신청

      0

      복사신청

      0

      EDDS신청

      0

      동일 주제 내 활용도 TOP

      더보기

      주제

      연도별 연구동향

      연도별 활용동향

      연관논문

      연구자 네트워크맵

      공동연구자 (7)

      유사연구자 (20) 활용도상위20명

      이 자료와 함께 이용한 RISS 자료

      나만을 위한 추천자료

      해외이동버튼