Radiation therapy using flattening filter free beam can prevent beam attenuation caused by flattening filter and can improve treatment efficiency. However, accurate dose control is not established for nonuniform iso dose distributions. In this study, ...
Radiation therapy using flattening filter free beam can prevent beam attenuation caused by flattening filter and can improve treatment efficiency. However, accurate dose control is not established for nonuniform iso dose distributions. In this study, curved dosimeter based on photoconductive material HgI<sub>2</sub> was fabricated and its reproducibility and linearity were evaluated at 6 MV photon energy to verify its performance. In order to show the usefulness of the curved measurement, the signals measured on the flat substrate and the curved substrate were compared in the flattening filter free beam using the acrylic filter. As a result, the reproducibility of the unit cell dosimeter was evaluated as SE 0.613%, and the linearity was evaluated as R-sq 0.9999. The usability evaluation of the array curve dosimeter demonstrated its usefulness by indicating a curvature error rate of 11.073%p and a reduced error rate.