      이온크로마토그래프를 이용한 수중의 퍼클로레이트 농축 및 분석 = Analysis of Perchlorate in Water Using Ion Chromatograph with Preconcentration

      한글로보기

      https://www.riss.kr/link?id=A105665034

      • 0

        상세조회
      • 0

        다운로드
      서지정보 열기
      • 내보내기
      • 내책장담기
      • 공유하기
      • 오류접수

      부가정보

      다국어 초록 (Multilingual Abstract)

      This study included the development of analytical method for determining perchlorate in water sample. The analytical condition was referred in EPA 314.0 method which use ion chromatography and the concentrator column was replaced by the guard column. Concentrating 10mL raw or treated water sample on to AGl6 guard column made it possible to get the LOD(Limit of Detection) of $0.73\;{\mu}g/L$. The total run time was 11 minutes and during run time next sample could be concentrated on AGl6 guard column. Compared to the Concentration method which needed manual operation, the Direct Injection method could screen the many water samples. The LOD of the Direct Injection method was higher and the sensitivity was lower than that of the Concentration method. The RSDs(Relative Standard Deviations) were lower than 2.5 % for peak height and 0.7 % for retention time in pre-concentration methods. This method Showed good reproducibility and reliability and it was thought the deviations of recovery value could be reduced by considering column capacity and making water sample homogeneous. Matrix Elimination could be done using the pre-concentration method if perchlorate were in complex matrix of sample.
      번역하기

      This study included the development of analytical method for determining perchlorate in water sample. The analytical condition was referred in EPA 314.0 method which use ion chromatography and the concentrator column was replaced by the guard column. ...

      This study included the development of analytical method for determining perchlorate in water sample. The analytical condition was referred in EPA 314.0 method which use ion chromatography and the concentrator column was replaced by the guard column. Concentrating 10mL raw or treated water sample on to AGl6 guard column made it possible to get the LOD(Limit of Detection) of $0.73\;{\mu}g/L$. The total run time was 11 minutes and during run time next sample could be concentrated on AGl6 guard column. Compared to the Concentration method which needed manual operation, the Direct Injection method could screen the many water samples. The LOD of the Direct Injection method was higher and the sensitivity was lower than that of the Concentration method. The RSDs(Relative Standard Deviations) were lower than 2.5 % for peak height and 0.7 % for retention time in pre-concentration methods. This method Showed good reproducibility and reliability and it was thought the deviations of recovery value could be reduced by considering column capacity and making water sample homogeneous. Matrix Elimination could be done using the pre-concentration method if perchlorate were in complex matrix of sample.

      더보기