      평판형 검출기를 이용한 디지털 방사선 촬영의 방사선량 감소에 따른 화질에 대한 영향: 가토를 이용한 실험적 연구 = The Effect of Dose Reduction on Image Quality in Digital Radiography Using a Flat-panel Detector: Experimental Study in Rabbits

      한글로보기

      https://www.riss.kr/link?id=A104533924

      • 0

        상세조회
      • 0

        다운로드
      서지정보 열기
      • 내보내기
      • 내책장담기
      • 공유하기
      • 오류접수

      부가정보

      다국어 초록 (Multilingual Abstract)

      Purpose: To evaluate the effect of dose reduction on image quality in digital radiography using a flat-panel detector.
      Materials and Methods: Digital radiographs of 30 rabbits were obtained at two different dose levels (33.23 μGy for the standard dose group and 20.09 μGy for the reduced dose group). The amorphous selenium-based flatpanel detector system had a panel size of 7×8.5 inches, a matrix of 1280×1536 (pixels?), and a pixel pitch of 138 μm. Four observers evaluated the soft-copy images on a high-resolution video monitor (2560×2048×8 bits) in random order. The observers rated the visibility of 13 different anatomic structures on a 5-point scale, viz. the retrocardiac lung, subdiaphragmatic lung, heart border, diaphragmatic border, proximal airway, unobscured lung, liver border, kidney border, bowel gas, flank stripe, ribs, and vertebrae in the mediastinal and abdominal regions. Statistical significance was determined using Wilcoxon’s signed rank test.
      Results: There was no statistically significant difference in the visibility of the anatomic structures on digital radiography between the standard and reduced dose groups.
      Conclusion: Digital radiography using an amorphous selenium-based flat-panel detector can preserve the image quality, even though the does is reduced to 40% of the standard level.
      번역하기

      Purpose: To evaluate the effect of dose reduction on image quality in digital radiography using a flat-panel detector. Materials and Methods: Digital radiographs of 30 rabbits were obtained at two different dose levels (33.23 μGy for the standard dos...

      Purpose: To evaluate the effect of dose reduction on image quality in digital radiography using a flat-panel detector.
      Materials and Methods: Digital radiographs of 30 rabbits were obtained at two different dose levels (33.23 μGy for the standard dose group and 20.09 μGy for the reduced dose group). The amorphous selenium-based flatpanel detector system had a panel size of 7×8.5 inches, a matrix of 1280×1536 (pixels?), and a pixel pitch of 138 μm. Four observers evaluated the soft-copy images on a high-resolution video monitor (2560×2048×8 bits) in random order. The observers rated the visibility of 13 different anatomic structures on a 5-point scale, viz. the retrocardiac lung, subdiaphragmatic lung, heart border, diaphragmatic border, proximal airway, unobscured lung, liver border, kidney border, bowel gas, flank stripe, ribs, and vertebrae in the mediastinal and abdominal regions. Statistical significance was determined using Wilcoxon’s signed rank test.
      Results: There was no statistically significant difference in the visibility of the anatomic structures on digital radiography between the standard and reduced dose groups.
      Conclusion: Digital radiography using an amorphous selenium-based flat-panel detector can preserve the image quality, even though the does is reduced to 40% of the standard level.

      더보기

      국문 초록 (Abstract)

      목적: 평판형 검출기를 이용한 디지털 방사선 촬영에서 방사선량 감소의 화질에 대한 평가를 하고자 한다.
      대상과 방법: 가토 30마리를 대상으로 표준군은 33.23 μGy, 감소군은 20.09 μGy으로2가지의 다른 방사선량으로 디지털 방사선촬영을 시행하였다. 사용된 평판형 비정질 셀레늄 검출기는 7×8.5 inch 크기와 1280×1536의 matrix, 138 μm의 pixel pitch로 구성되었다. 획득된 2군의 영상을 고해상도 모니터에서 4명의 판독자가 심장후 폐야, 횡격막후 폐야, 심장 윤곽, 횡격막 윤곽, 근위부 기도, 가리지 않은 폐야 등의 흉부 구조물, 간 윤곽, 좌우측 신장 윤곽, 장내 가스, 외측 복벽 지방선 등의 복부 구조물, 늑골, 종격동뒤 척추, 복강부위 척추 등의 골 구조물 등 총 13가지의 해부학적 구조물에 대하여 5가지 척도로 평가하였고 결과에 대하여 Wilcoxon’s signed ranks test를 이용하여 통계적 처리를 하였다.
      결과: 표준군과 감소군간에 13 가지 해부학적 구조물에 대하여 영상의 질적 평가상 통계적으로 유의한 차이는 없었다.
      결론: 평판형 비정질 셀레늄 검출기를 사용한 디지털 방사선촬영술은 기존의 방사선량에서 40%의 감소를 하여도 영상의 질을 유지할 수 있었다.
      번역하기

      목적: 평판형 검출기를 이용한 디지털 방사선 촬영에서 방사선량 감소의 화질에 대한 평가를 하고자 한다. 대상과 방법: 가토 30마리를 대상으로 표준군은 33.23 μGy, 감소군은 20.09 μGy으로2가�...

      목적: 평판형 검출기를 이용한 디지털 방사선 촬영에서 방사선량 감소의 화질에 대한 평가를 하고자 한다.
      대상과 방법: 가토 30마리를 대상으로 표준군은 33.23 μGy, 감소군은 20.09 μGy으로2가지의 다른 방사선량으로 디지털 방사선촬영을 시행하였다. 사용된 평판형 비정질 셀레늄 검출기는 7×8.5 inch 크기와 1280×1536의 matrix, 138 μm의 pixel pitch로 구성되었다. 획득된 2군의 영상을 고해상도 모니터에서 4명의 판독자가 심장후 폐야, 횡격막후 폐야, 심장 윤곽, 횡격막 윤곽, 근위부 기도, 가리지 않은 폐야 등의 흉부 구조물, 간 윤곽, 좌우측 신장 윤곽, 장내 가스, 외측 복벽 지방선 등의 복부 구조물, 늑골, 종격동뒤 척추, 복강부위 척추 등의 골 구조물 등 총 13가지의 해부학적 구조물에 대하여 5가지 척도로 평가하였고 결과에 대하여 Wilcoxon’s signed ranks test를 이용하여 통계적 처리를 하였다.
      결과: 표준군과 감소군간에 13 가지 해부학적 구조물에 대하여 영상의 질적 평가상 통계적으로 유의한 차이는 없었다.
      결론: 평판형 비정질 셀레늄 검출기를 사용한 디지털 방사선촬영술은 기존의 방사선량에서 40%의 감소를 하여도 영상의 질을 유지할 수 있었다.

      더보기

      참고문헌 (Reference)

      1 Chotas HG, "Technical evaluation of a digi-tal chest radiography system that uses a selenium detector" 195 : 264-270, 1995

      2 Willis CE, "Strategies for dose reduction in ordinary radiographicexaminations using CR and DR" 34 (34): 196-200, 2004

      3 Ludwig K, "Selenium-based digital radiography in the detection ofbone lesions:preliminary experience with experimentally createddefects" 216 : 220-224, 2000

      4 Schaefer-Prokop CM, "Selenium radiography versus storage phosphor and conven-tional radiography in the detection of simulated chest lesions" 201 : 45-50, 1996

      5 Faulkner K, "Radiation dose to neonates on aspecial care baby unit" 62 : 230-233, Radiol1989

      6 Chotas HG, "Principles of digital radiog-raphy with large-area, electronically readable detectors: a reviewof the basics." 210 : 595-599, 1999

      7 Granfors PR, "Performance of a 41×41 cm2 amor-phous silicon flat panel X-ray detector for radiographic imaging ap-plications" 27 : 1324-1331, 2000

      8 Jones NF, "Neonatal chest and abdominal ra-diation dosimetry:a comparison of two radiographic techniques" 74 : 920-925, 2001

      9 Jabri KN, "Management of pediatric radiationdose using GE ’s revolution digital radiography systems" 34 (34): 215-220, 2004

      10 National RadiologicalProtection Board, "Institute of Physical Sciences in Medicine College of Radiographers National Protocol forPatient Dose Measurements in Diagnostic Radiology" 1992

      1 Chotas HG, "Technical evaluation of a digi-tal chest radiography system that uses a selenium detector" 195 : 264-270, 1995

      2 Willis CE, "Strategies for dose reduction in ordinary radiographicexaminations using CR and DR" 34 (34): 196-200, 2004

      3 Ludwig K, "Selenium-based digital radiography in the detection ofbone lesions:preliminary experience with experimentally createddefects" 216 : 220-224, 2000

      4 Schaefer-Prokop CM, "Selenium radiography versus storage phosphor and conven-tional radiography in the detection of simulated chest lesions" 201 : 45-50, 1996

      5 Faulkner K, "Radiation dose to neonates on aspecial care baby unit" 62 : 230-233, Radiol1989

      6 Chotas HG, "Principles of digital radiog-raphy with large-area, electronically readable detectors: a reviewof the basics." 210 : 595-599, 1999

      7 Granfors PR, "Performance of a 41×41 cm2 amor-phous silicon flat panel X-ray detector for radiographic imaging ap-plications" 27 : 1324-1331, 2000

      8 Jones NF, "Neonatal chest and abdominal ra-diation dosimetry:a comparison of two radiographic techniques" 74 : 920-925, 2001

      9 Jabri KN, "Management of pediatric radiationdose using GE ’s revolution digital radiography systems" 34 (34): 215-220, 2004

      10 National RadiologicalProtection Board, "Institute of Physical Sciences in Medicine College of Radiographers National Protocol forPatient Dose Measurements in Diagnostic Radiology" 1992

      11 Council of the European Communities, "European guidelines on qual-ity criteria for diagnostic radiographic images in pediatrics" 1996

      12 Goo JM, "Digitalchest radiography with a selenium-based flat-panel detector versusa storage phosphor system:comparison of soft-copy images" 175 : 1013-1018, 2000

      13 Volk M, "Digital radiography of the skeleton using alarge-area detector based on amorphous silicon technology:imagequality and potential for dose reduction in comparison withscreen-film radiography" 55 : 615-621, 2000

      14 Van Heesewijk HP, "Digital chest imaging with a selenium detector:com-parison with conventional radiography for visualization of specificanatomic regions of the chest" 165 : 535-540, 1995

      15 Antonuk LE, "Demonstration of megavoltage and diagnostic x-rayimaging with hydrogenated amorphous silicon arrays" 19 : 1455-1466, Phys1992

      16 Fink C, "Clinical comparative study with a large-areaamorphous silicon flat-panel detector:image quality and visibilityof anatomic structures on chest radiography" 178 : 481-486, 2002

      17 Van Heesewijk HP, "Chest imaging with a selenium detector versus conventionalfilm radiography:a CT-controlled study" 200 : 687-690, 1996

      18 Launders JH, "A comprehensive physical image quality evaluationof a selenium-based digital x-ray imaging system for thorax radiog-raphy" 25 : 986-997, 1998

      더보기

      분석정보

      View

      상세정보조회

      0

      Usage

      원문다운로드

      0

      대출신청

      0

      복사신청

      0

      EDDS신청

      0

      동일 주제 내 활용도 TOP

      더보기

      주제

      연도별 연구동향

      연도별 활용동향

      연관논문

      연구자 네트워크맵

      공동연구자 (7)

      유사연구자 (20) 활용도상위20명

      인용정보 인용지수 설명보기

      학술지 이력

      학술지 이력
      연월일 이력구분 이력상세 등재구분
      2024 평가예정 해외DB학술지평가 신청대상 (해외등재 학술지 평가)
      2021-01-01 평가 등재학술지 유지 (해외등재 학술지 평가) KCI등재
      2020-01-01 평가 등재학술지 유지 (재인증) KCI등재
      2017-01-01 평가 등재학술지 유지 (계속평가) KCI등재
      2016-11-24 학술지명변경 외국어명 : Journal of The Korean Radiological Society -> Journal of the Korean Society of Radiology (JKSR) KCI등재
      2016-11-15 학회명변경 영문명 : The Korean Radiological Society -> The Korean Society of Radiology KCI등재
      2013-01-01 평가 등재 1차 FAIL (등재유지) KCI등재
      2010-01-01 평가 등재학술지 유지 (등재유지) KCI등재
      2008-01-01 평가 등재학술지 유지 (등재유지) KCI등재
      2006-01-01 평가 등재학술지 유지 (등재유지) KCI등재
      2005-09-15 학술지명변경 한글명 : 대한방사선의학회지 -> 대한영상의학회지 KCI등재
      2003-01-01 평가 등재학술지 선정 (등재후보2차) KCI등재
      2002-01-01 평가 등재후보 1차 PASS (등재후보1차) KCI등재후보
      2000-07-01 평가 등재후보학술지 선정 (신규평가) KCI등재후보
      더보기

      학술지 인용정보

      학술지 인용정보
      기준연도 WOS-KCI 통합IF(2년) KCIF(2년) KCIF(3년)
      2016 0.1 0.1 0.07
      KCIF(4년) KCIF(5년) 중심성지수(3년) 즉시성지수
      0.06 0.05 0.258 0.01
      더보기

      이 자료와 함께 이용한 RISS 자료

      나만을 위한 추천자료

      해외이동버튼