Line spread function(LSF) in computed radiography was measured by an program, testing for measurement digital radiograph, to measure density automatically.
The results were as follows : the width between two lines in slit assembly have to be reappear...
Line spread function(LSF) in computed radiography was measured by an program, testing for measurement digital radiograph, to measure density automatically.
The results were as follows : the width between two lines in slit assembly have to be reappeared black lines with 10㎛ on IP image, but that was measured hundreds of microns because of diffusion of light in phosphor.
LSF on digital image could be measured by new computerized profile, therefore, capability of digital measurement is also expected on modulation transfer function(MTF) for measuring analog signal.