      전산화단층촬영 단말장치와 개인용 컴퓨터에서 재구성한 두부 3차원 전산화단층영상의 비교 = Comparison of personal computer with CT workstation in the evaluation of 3-dimensional CT image of the skull

      한글로보기

      https://www.riss.kr/link?id=A3009505

      • 0

        상세조회
      • 0

        다운로드
      서지정보 열기
      • 내보내기
      • 내책장담기
      • 공유하기
      • 오류접수

      부가정보

      다국어 초록 (Multilingual Abstract) kakao i 다국어 번역

      Purpose : To evaluate the usefulness of the reconstructed 3-dimensional image on the personal computer in comparison with that of the CT workstation by quantitative comparison and analysis.
      Materials and Methods : The spiral CT data obtained from 27 persons were transferred from the CT workstation to a personal computer, and they were reconstructed as 3-dimensional image on the personal computer using V-works 2.OTM. One observer obtained the 14 measurements on the reconstructed 3-dimensional image on both the CT workstation and the personal computer. Paired t-test was used to evaluate the intraobserver difference and the mean value of the each measurement on the CT workstation and the personal computer. Pearson correlation analysis and % incongruence were also performed.
      Results : I-Gn, N-Gn, N-A, N-Ns, B-A, and G-Op did not show any statistically significant difference (p)0.05), B-O, B-N, Eu-Eu, Zy-Zy, Biw, D-D, Orbrd R, and L had statistically significant difference(p〈0.05), but the mean values of the differences of all measurements were below 2 mm, except for D-D. The value of correlation coefficient γ was greater than 0.95 at I-Gn, N-Gn, N-A, N-Ns, B-A, B-N, G-Op, Eu-Eu, Zy-Zy, and Biw, and it was 0.75 at B-O, 0.78 at D-D, and 0.82 at both Orbrd R and L. The % incongruence was below 4% at I-Gn, N-Gn, N-A, N-Ns, B-A, B-N, G-Op, Eu-Eu, Zy-Zy, and Biw, and 7.18%, 10.78%, 4.97%, 5.89% at B-O, D-D, Orbrd R and L respectively.
      Conclusion : It can be considered that the utilization of the personal computer has great usefulness in reconstruction of the 3-dimensional image when it comes to the economics, accessibility and convenience, except for thin bones and the landmarks which are difficult to be located.(Korean J Oral Maxillofac Radiol 2001;31:1-7)
      번역하기

      Purpose : To evaluate the usefulness of the reconstructed 3-dimensional image on the personal computer in comparison with that of the CT workstation by quantitative comparison and analysis. Materials and Methods : The spiral CT data obtained from 27 ...

      Purpose : To evaluate the usefulness of the reconstructed 3-dimensional image on the personal computer in comparison with that of the CT workstation by quantitative comparison and analysis.
      Materials and Methods : The spiral CT data obtained from 27 persons were transferred from the CT workstation to a personal computer, and they were reconstructed as 3-dimensional image on the personal computer using V-works 2.OTM. One observer obtained the 14 measurements on the reconstructed 3-dimensional image on both the CT workstation and the personal computer. Paired t-test was used to evaluate the intraobserver difference and the mean value of the each measurement on the CT workstation and the personal computer. Pearson correlation analysis and % incongruence were also performed.
      Results : I-Gn, N-Gn, N-A, N-Ns, B-A, and G-Op did not show any statistically significant difference (p)0.05), B-O, B-N, Eu-Eu, Zy-Zy, Biw, D-D, Orbrd R, and L had statistically significant difference(p〈0.05), but the mean values of the differences of all measurements were below 2 mm, except for D-D. The value of correlation coefficient γ was greater than 0.95 at I-Gn, N-Gn, N-A, N-Ns, B-A, B-N, G-Op, Eu-Eu, Zy-Zy, and Biw, and it was 0.75 at B-O, 0.78 at D-D, and 0.82 at both Orbrd R and L. The % incongruence was below 4% at I-Gn, N-Gn, N-A, N-Ns, B-A, B-N, G-Op, Eu-Eu, Zy-Zy, and Biw, and 7.18%, 10.78%, 4.97%, 5.89% at B-O, D-D, Orbrd R and L respectively.
      Conclusion : It can be considered that the utilization of the personal computer has great usefulness in reconstruction of the 3-dimensional image when it comes to the economics, accessibility and convenience, except for thin bones and the landmarks which are difficult to be located.(Korean J Oral Maxillofac Radiol 2001;31:1-7)

      더보기

      목차 (Table of Contents)

      • 서론
      • 재료 및 방법
      • 1.연구재료
      • 2.연구방법
      • 연구결과
      • 서론
      • 재료 및 방법
      • 1.연구재료
      • 2.연구방법
      • 연구결과
      • 총괄 및 고찰
      더보기

      동일학술지(권/호) 다른 논문

      동일학술지 더보기

      더보기

      분석정보

      View

      상세정보조회

      0

      Usage

      원문다운로드

      0

      대출신청

      0

      복사신청

      0

      EDDS신청

      0

      동일 주제 내 활용도 TOP

      더보기

      주제

      연도별 연구동향

      연도별 활용동향

      연관논문

      연구자 네트워크맵

      공동연구자 (7)

      유사연구자 (20) 활용도상위20명

      이 자료와 함께 이용한 RISS 자료

      나만을 위한 추천자료

      해외이동버튼