THMs and PCBs were separated and analysed with elution on Novapak ODS or $\mu-Bondapak$ phenyl column by an eluent containing p-nitrophenol (p-NP). THMs studied were CHCl3, CHBrCl2, CHBr2Cl, and CHBr3, and PCBs used were Aroclor 1221, 1242, 1248, $\al...
THMs and PCBs were separated and analysed with elution on Novapak ODS or $\mu-Bondapak$ phenyl column by an eluent containing p-nitrophenol (p-NP). THMs studied were CHCl3, CHBrCl2, CHBr2Cl, and CHBr3, and PCBs used were Aroclor 1221, 1242, 1248, $\alpha-$ and $\beta-BHC.$ It was thought that the retention on the stationary phase and sensitivities of the samples are related to the interaction between the sample and stationary phase or p-NP. THMs were separated completely on the ODS column by elution with MeOH-water (30 : 70) containing $1.0{\times}10^{-4}$ M p-NP and some of PCBs were separated on the phenyl column by elution with $CH_3CN$-water(50 : 50) containing $1.0{\times}10^{-4}$ M p-NP. Detection limits of THMs were from $1.0{\times}10^{-4}$ g to $1.0{\times}10^{-6}$ g. Aroclors were $2{\times}10^{-6}$ g, and $\alpha-$ and $\beta-BHC$ were $2{\times}10^{-4}$ g and $1.0{\times}10^{-4}$ g respectively.