Triple points of high purity materials have been used to calibrate primarily the capsule-type platinum resistance thermometer (PRT) in the temperature range of the triple point of equilibrium hydrogen (13.8033 K) and water (273.16 K). In this work, tr...
Triple points of high purity materials have been used to calibrate primarily the capsule-type platinum resistance thermometer (PRT) in the temperature range of the triple point of equilibrium hydrogen (13.8033 K) and water (273.16 K). In this work, triple points of Ne, O₂, Ar, Hg and Hz0 except for the triple point of equilibrium hydrogen were realized to establish the International Temperature Scale of 1990 (ITS-90). At each fixed point, two capsule-type PRTs, which were selected for the international comparison, were tested two times. The combined uncertainties of the realization of each triple point were calculated considering the type A and type B evaluation. In Korea Research Institute of Standards and Science, the combined standard uncertainties of the defining triple points by the ITS-90 were estimated to about 0.18 mK for Ne, 0.14 mK for O₂, 0.14 mK for Ar, 0.24 mK for Hg and 0.11 mK for H₂O, respectively.
마이크로 열 센서용 측온저항체 온도센서의 제작 및 특성
변위센서응용을 위한 피라미드형 실리콘 턴널링소자의 제조
표면 미세 가공된 측면형 전계 방출 소자를 이용한 초소형 진공 센서의 제작