In order to assay the human plasma concentration of nifedipine in patients with bronchopulmonary dysplasia (BPD) and pulmonary hypertension, a modified high performance liquid chromatography (HPLC) method was applied. The retention times for nifedipin...
In order to assay the human plasma concentration of nifedipine in patients with bronchopulmonary dysplasia (BPD) and pulmonary hypertension, a modified high performance liquid chromatography (HPLC) method was applied. The retention times for nifedipine and an internal standard (11-ketoprogesterone) were 10.5 ± 0.41 and 13.1 ± 0.63 min, respectively. Absolute recovery from the plasma was 102.9 ± 7.07%. Reproducibility was excellent and variability between the runs was small. There was a negligible degradation during the assay procedure. The calibration curve shows a good linearity in the range of the desired plasma concentrations of nifedipine. A stability test of nifedipine in the human plasma shows 8 and 13% degradation during the storage of 5 and 9 months, respectively. There were no interferences on the HPLC assay with any possible medications for the BPD. The method has been used to monitor the drug concentrations in a patient. The concentration-time curve of a patient after a single oral dose of 0.3 ㎎/㎏ shows a double-peak phenomenon that was quite different from the previous report, suggesting non-bolus administration. However the hemodynamic responses were corresponding to the plasma concentration levels of nifedipine.