    고주파 소작기의 전극 디자인에 따른 출력 특성 연구 Ⅱ = A Study on the Output Characteristics of High-Frequency Ablation Devices based on Electrode Design II

    한글로보기

    https://www.riss.kr/link?id=A109895125

    • 0

      상세조회
    • 0

      다운로드
    서지정보 열기
    • 내보내기
    • 내책장담기
    • 공유하기
    • 오류접수
    인용문이 복사되었습니다.

    부가정보

    다국어 초록 (Multilingual Abstract) kakao i 다국어 번역

    Five novel miniature bipolar radiofrequency (RF) electrode tips with distinct tip geometries (spherical, flat, square, and 45° angled) were developed to enable high-precision tissue ablation. Performance was evaluated on saline-soaked tissue, ex vivo bovine liver, and porcine muscle under consistent RF power settings. All designs produced highly localized lesions only a few millimeters across, confirming precise ablation with minimal damage to surrounding tissue. Tip geometry influenced ablation efficiency: a 45° angled tip created ~5 mm lesions at lower power (highest efficiency), whereas an ultra-fine 1.0 mm tip produced ~1 mm lesions but required higher power. These results indicate that the new bipolar RF electrodes achieve precise, localized tissue ablation with minimal surrounding tissue damage and show promise for precise lesion removal in minimally invasive surgery.
    번역하기

    Five novel miniature bipolar radiofrequency (RF) electrode tips with distinct tip geometries (spherical, flat, square, and 45° angled) were developed to enable high-precision tissue ablation. Performance was evaluated on saline-soaked tissue, ex vivo...

    Five novel miniature bipolar radiofrequency (RF) electrode tips with distinct tip geometries (spherical, flat, square, and 45° angled) were developed to enable high-precision tissue ablation. Performance was evaluated on saline-soaked tissue, ex vivo bovine liver, and porcine muscle under consistent RF power settings. All designs produced highly localized lesions only a few millimeters across, confirming precise ablation with minimal damage to surrounding tissue. Tip geometry influenced ablation efficiency: a 45° angled tip created ~5 mm lesions at lower power (highest efficiency), whereas an ultra-fine 1.0 mm tip produced ~1 mm lesions but required higher power. These results indicate that the new bipolar RF electrodes achieve precise, localized tissue ablation with minimal surrounding tissue damage and show promise for precise lesion removal in minimally invasive surgery.