In this paper, an implantable middle ear hearing aids using an electromagnetic transducer which consists of a tiny coil and a magnet is designed, and objective design method to implement a vibrator of the electromagnetic type transducer is proposed. S...
In this paper, an implantable middle ear hearing aids using an electromagnetic transducer which consists of a tiny coil and a magnet is designed, and objective design method to implement a vibrator of the electromagnetic type transducer is proposed. Sound characteristic of the electromagnetic type vibrator is excellent but power transfer rate is lower than that of a ceramic type. To improve power transfer efficiency of system, external and inner part of a middle ear system is designed using FI\T transmission method. We implemented an experimental set of the designed transducer, inner part, and outer part. Then, we measured vibration characteristics of a metal strip and a temporal bone of a dead body. As the results, we confirmed the advantage of FM method and showed that designed transducer could effectively transmit vibration of amplified input sound pressure to ossicle.