Scatter correction for single photon emission computed tomography (SPECT) plays an important role to improve image quality and quantitation. The purpose of this study was to investigate three scatter correction methods using Monte Carlo simulation. Point source and Jaszack phantom filled with Tc-99m were simulated by Monte Carlo code, SIMIND. For scatter correction, we applied three methods, Compton window (CW) method, triple window (TW) method, and dual photopeak window (DPW) method. Point sources located at various depths along the center line within a 20-cm phantom were simulated to calculate the window ratios and corresponding scatter fractions by evaluating the polynomial coefficients for DPW method. Energy windows were located in W$_1$=92-125 keV, W$_2$=124-126 keV, W$_3$=136-140 keV, W$_4$=140-141 keV, and W$_{5}$=154-156 keV. The results showed that in Jaszack phantom with cold sphere and hot sphere, the TW gave the closest contrast and percentage recovery to the ideal image, respectively, while CW overestimated and DPW underestimated the contrast of ideal one. All three scatter correction methods showed an improved image contrast. In conclusion, scatter correction is essential for improving image contrast and accurate quantification. The choice of scatter correction method should be made on the basis of accuracies and ease of implementation.