      FIT OF FIXTURE/ABUTMENT/SCREW INTERFACES OF INTERNAL CONNECTION IMPLANT SYSTEM

      한글로보기

      https://www.riss.kr/link?id=A45026491

      • 0

        상세조회
      • 0

        다운로드
      서지정보 열기
      • 내보내기
      • 내책장담기
      • 공유하기
      • 오류접수

      부가정보

      다국어 초록 (Multilingual Abstract)

      Statement of problem. Accurate fit between the implant components is important because the misfit of the implant components results in frequent screw loosening, irreversible screw fracture, plaque accumulation, poor soft tissue reaction, and destruction of osseointegration.
      Purpose. This study is to evaluate the machining accuracy and consistency of the implant fixture/abutment/screw interfaces of the internal connection system by using a Stereoscopic Zoom microscope and FE-SEM(field emission scanning electron microscope)
      Materials and methods. The implant systems selected in this study were internal connection type implants from AVANA(Osstem^(?)), Bioplant(Cowell-Medi^(?)), Dio(DIO^(?)), Neoplant(Neobiotech^(?)), Implantium(Dentium^(?))systems. Each group was acquired 2 fixtures at random. Two piece type abutment and one piece type abutment for use with each implant System were acquired. Screw were respectively used to hold a two piece type abutment to a implant fixture.
      The implant fixtures were perpendiculary mounted in acrylic resin block. Each two piece abutment was secured to the implant fixture by screw and one piece abutment also secured to the implant fixture. Abutment/fixture assembly were mounted in liquid unsaturated polyester. All samples were cross-sectioned with grinder-polisher unit. Finally all specimens were analysed the fit between implant fixture/abutment/screw interfaces
      Results and conclusions.
      1. Implant fixture/abutment/screw connection interfaces of internal connection systems made in Korea were in good condition.
      2. The results of the above study showed that materials and mechanical properties and quality of milling differed depending an their manufacturing companies.
      번역하기

      Statement of problem. Accurate fit between the implant components is important because the misfit of the implant components results in frequent screw loosening, irreversible screw fracture, plaque accumulation, poor soft tissue reaction, and destructi...

      Statement of problem. Accurate fit between the implant components is important because the misfit of the implant components results in frequent screw loosening, irreversible screw fracture, plaque accumulation, poor soft tissue reaction, and destruction of osseointegration.
      Purpose. This study is to evaluate the machining accuracy and consistency of the implant fixture/abutment/screw interfaces of the internal connection system by using a Stereoscopic Zoom microscope and FE-SEM(field emission scanning electron microscope)
      Materials and methods. The implant systems selected in this study were internal connection type implants from AVANA(Osstem^(?)), Bioplant(Cowell-Medi^(?)), Dio(DIO^(?)), Neoplant(Neobiotech^(?)), Implantium(Dentium^(?))systems. Each group was acquired 2 fixtures at random. Two piece type abutment and one piece type abutment for use with each implant System were acquired. Screw were respectively used to hold a two piece type abutment to a implant fixture.
      The implant fixtures were perpendiculary mounted in acrylic resin block. Each two piece abutment was secured to the implant fixture by screw and one piece abutment also secured to the implant fixture. Abutment/fixture assembly were mounted in liquid unsaturated polyester. All samples were cross-sectioned with grinder-polisher unit. Finally all specimens were analysed the fit between implant fixture/abutment/screw interfaces
      Results and conclusions.
      1. Implant fixture/abutment/screw connection interfaces of internal connection systems made in Korea were in good condition.
      2. The results of the above study showed that materials and mechanical properties and quality of milling differed depending an their manufacturing companies.