The ${\gamma}$-ray deposition spectra were calculated by Monte Calro method to obtain the scintillation characteristics of the ${\gamma}$-ray for BGO scintillation detector with the spherical shape of 1.25 cm radius. The code used in calculating the $...
The ${\gamma}$-ray deposition spectra were calculated by Monte Calro method to obtain the scintillation characteristics of the ${\gamma}$-ray for BGO scintillation detector with the spherical shape of 1.25 cm radius. The code used in calculating the ${\gamma}$-ray deposition spectra was made for personal computer with qbasic language. Also the ${\gamma}$-ray energy spectra of $^{22}Na$, $^{137}Cs$ and $^{207}Bi$ were measured with the detector. The energy dependent resolution below 2000 keV for the detector was determined by estimating the standard deviation of the photopeak fitted with gaussian function, and $X^{2}$ fitting using Nardi's empirical formula. The measured spectra of $^{22}Na$ and $^{137}Cs$ were compared with the broadened spectra which were obtained by broadening the calculated ${\gamma}$-ray deposition spectra with the energy dependent resolution. The absolute efficiency and the intrinsic peak efficiency of the detector were obtained by calculating the ${\gamma}$-ray deposition spectrum with the code.