We realized the national standard of the candela, one of the SI units, by using two photometers with the spectral responsivity measured in reference to the absolute cryogenic radiometer. The external apertures of the photometers were fabricated using ...
We realized the national standard of the candela, one of the SI units, by using two photometers with the spectral responsivity measured in reference to the absolute cryogenic radiometer. The external apertures of the photometers were fabricated using a diamond turning machine, and measured in terms of area with uncertainty of 0.05 %(k = 1). The candela is realized using a 1 kW FEL lamp and the characterized photometers on an optical bench. The uncertainty is budgeted to be 0.25 %(k = 1) considering the uncertainty of the spectral responsivity and the response uniformity of the detectors, the area of the external apertures, the color temperature of the lamp, and the positioning reproducibility of the photometers and the lamp. We verified the realized scale by comparing with the scale of National Institute of Standards and Technology, USA. They coincided with each other within 0.1%.