    The ISFET glucose and sucrose sensors containing platinum electrode and photopolymeric enzyme membrane were fabricated. The platinum working electrode was used for the electrolysis of hydrogen peroxide, which was the other product of the enzyme reaction, to improve sensing characteristics of the sensors. In order to improve response time, photo-crosslinkable polymer(PVA-SbQ) was used to the matrix for the enzyme immobilized membrane. The characteristics of glucose and sucrose sensors were investigated according to the variation of platinum electrode area. The response time was about $3{\sim}5$ minutes and determinations of glucose and sucrose in the range of about $30{\sim}300mg/dl$ could be possible.
    번역하기

    The ISFET glucose and sucrose sensors containing platinum electrode and photopolymeric enzyme membrane were fabricated. The platinum working electrode was used for the electrolysis of hydrogen peroxide, which was the other product of the enzyme reacti...

    The ISFET glucose and sucrose sensors containing platinum electrode and photopolymeric enzyme membrane were fabricated. The platinum working electrode was used for the electrolysis of hydrogen peroxide, which was the other product of the enzyme reaction, to improve sensing characteristics of the sensors. In order to improve response time, photo-crosslinkable polymer(PVA-SbQ) was used to the matrix for the enzyme immobilized membrane. The characteristics of glucose and sucrose sensors were investigated according to the variation of platinum electrode area. The response time was about $3{\sim}5$ minutes and determinations of glucose and sucrose in the range of about $30{\sim}300mg/dl$ could be possible.