      유방암 환자의 수술 중 방사선치료 시 피부선량 측정을 통한 안전성 평가 = Evaluation of safety by skin dosimetry in Intraoperative Radiotherapy for breast cancer patients

      한글로보기

      https://www.riss.kr/link?id=A101423630

      • 0

        상세조회
      • 0

        다운로드
      서지정보 열기
      • 내보내기
      • 내책장담기
      • 공유하기
      • 오류접수

      부가정보

      다국어 초록 (Multilingual Abstract) kakao i 다국어 번역

      Purpose : We investigated the safety of Intrabeam$^{TM}$ system, X-ray unit for Intraoperative Radiotheray (IORT) by measuring surface dose using Optically Stimulated Luminescent Dosimeter(OSLD). Materials and Methods : 30 patients were selected, who were in breast cancer patients and had an operation of breast conserving surgery (BCS). At the inner surface of tumor bed, 20 Gy were described, and 5 Gy at 1cm depth from the inner surface. Along the size of tumor bed which could be decided after resection of tumor, the size of applicator were determined. Usual treatment time were from 18 to 40 minutes. For the measurement of surface doses, OSLD were placed at superior(U1,2), inferior(D1,2), lateral(L1,2) and medial(M1,2) directions from the center of applicator. Each direction, two OSLD were placed at 0.5 cm and 1.5 cm from the center. Mean, maximum, and minimum doses were analyzed to be compared. Results : Mean values were U1 $2.23{\pm}0.80Gy$, U2 $1.54{\pm}0.53Gy$, D1 $1.73{\pm}0.63Gy$, D2 $1.25{\pm}0.45Gy$, L1 $1.95{\pm}0.82Gy$, L2 $1.38{\pm}0.42Gy$, M1 $2.03{\pm}0.70Gy$, and M2 $1.51{\pm}0.58Gy$. Maximum values were 4.34 Gy at U1, and Minimum values were 0.45 Gy at M2. 13.3 % of patient (4pts out of 30) were reported that surface dose were over 4 Gy. Conclusion : The fact that skin dose of all patients were less than 5 Gy based on OSLD measurement showed the safety of Intrabeam$^{TM}$ system. In the relatively small breast volume, the tendency that surface dose was increased had been shown, which was analyzed by the data of patients who irradiated over 4Gy at skin surface. Therefore, for appropriate indication for IORT, it is suggested that breast volume as well as the size and position of tumor should be carfully considered.
      번역하기

      Purpose : We investigated the safety of Intrabeam$^{TM}$ system, X-ray unit for Intraoperative Radiotheray (IORT) by measuring surface dose using Optically Stimulated Luminescent Dosimeter(OSLD). Materials and Methods : 30 patients were selected, who ...

      Purpose : We investigated the safety of Intrabeam$^{TM}$ system, X-ray unit for Intraoperative Radiotheray (IORT) by measuring surface dose using Optically Stimulated Luminescent Dosimeter(OSLD). Materials and Methods : 30 patients were selected, who were in breast cancer patients and had an operation of breast conserving surgery (BCS). At the inner surface of tumor bed, 20 Gy were described, and 5 Gy at 1cm depth from the inner surface. Along the size of tumor bed which could be decided after resection of tumor, the size of applicator were determined. Usual treatment time were from 18 to 40 minutes. For the measurement of surface doses, OSLD were placed at superior(U1,2), inferior(D1,2), lateral(L1,2) and medial(M1,2) directions from the center of applicator. Each direction, two OSLD were placed at 0.5 cm and 1.5 cm from the center. Mean, maximum, and minimum doses were analyzed to be compared. Results : Mean values were U1 $2.23{\pm}0.80Gy$, U2 $1.54{\pm}0.53Gy$, D1 $1.73{\pm}0.63Gy$, D2 $1.25{\pm}0.45Gy$, L1 $1.95{\pm}0.82Gy$, L2 $1.38{\pm}0.42Gy$, M1 $2.03{\pm}0.70Gy$, and M2 $1.51{\pm}0.58Gy$. Maximum values were 4.34 Gy at U1, and Minimum values were 0.45 Gy at M2. 13.3 % of patient (4pts out of 30) were reported that surface dose were over 4 Gy. Conclusion : The fact that skin dose of all patients were less than 5 Gy based on OSLD measurement showed the safety of Intrabeam$^{TM}$ system. In the relatively small breast volume, the tendency that surface dose was increased had been shown, which was analyzed by the data of patients who irradiated over 4Gy at skin surface. Therefore, for appropriate indication for IORT, it is suggested that breast volume as well as the size and position of tumor should be carfully considered.

      더보기

      분석정보

      View

      상세정보조회

      0

      Usage

      원문다운로드

      0

      대출신청

      0

      복사신청

      0

      EDDS신청

      0

      동일 주제 내 활용도 TOP

      더보기

      주제

      연도별 연구동향

      연도별 활용동향

      연관논문

      연구자 네트워크맵

      공동연구자 (7)

      유사연구자 (20) 활용도상위20명

      이 자료와 함께 이용한 RISS 자료

      나만을 위한 추천자료

      해외이동버튼