Ion-selective electrode responsive to the thiocynate ion prepared by using the quaternary ammonium salts as a active material and PVC as a membrane matrix. The effect of chemical structure and composition of active material, and the membrane thickness...
Ion-selective electrode responsive to the thiocynate ion prepared by using the quaternary ammonium salts as a active material and PVC as a membrane matrix. The effect of chemical structure and composition of active material, and the membrane thickness on the linear response. the detection limit, and Nernstian slope of the electrode studied. Under the above optimum conditions of membrane, the effect of pH and the selectivity coefficients to various interfering anions were compared and investigated.
It was concluded that the functions of thiocynate ion-selective electrode(ISE) were closely related to the chemical structure of the quaternary ammonium salts. The linear response, and the detection limit of the electrode potential increased with the increase of the carbon chain length of the alkyl group in the quaternary ammonium salts in the ascending order of Aliquat 330T, TOAT, TDAT, and TDDAT.
The optimum membrane thickness was 0.3㎜. The electrode characteristics was better with the decrease of the concentration of active material, and the best concentration was 3 weight percent.
The membrane potential was independent of the pH variation in the region from pH 2 to 12. The order of the selectivity coefficients is as follows : ClO₄^->I^->NO₃^->Br^->F^->Cl^->OA_0◎SO₄^(2-).