We developed a microfluidic platform able to control the trap and release of magnetic beads used for separation of a specific biomolecules. The magnetic beads can be trapped and released conditionally by controlling the difference between the Stokes f...
We developed a microfluidic platform able to control the trap and release of magnetic beads used for separation of a specific biomolecules. The magnetic beads can be trapped and released conditionally by controlling the difference between the Stokes force induced by the fluid flow and magnetic force resulting from a permanent magnet. The permanent magnet of CoNiP alloy is electroplated. It is characterized to have the 1369 Oe of coercivity, 1762 Gauss of remanence, and 0.603MGOe of (BH)max. Through the experimental and numerical investigation, the magnetic beads are trapped under the flow velocity of 17 ${\mu}m/s$ and are released perfectly above the velocity of 174 ${\mu}m/s$.