Zhiheng Huang,Chunchuan Gu,Jiajun Wen,Langlang Zhu 성균관대학교(자연과학캠퍼스) 성균나노과학기술원 2019 NANO Vol.14 No.1
In this paper, a new and one-pot electrodeposition method was expanded for the preparation of NiS nanoparticles-based electrochemical biosensor using metal-ion complexes as a precursor. Thioacetamide was used to control the production rate of NiS nanoparticles for the first time. The proposed electrochemical sensor was characterized by energy dispersive X-ray spectroscopy (EDX), field emission scanning electron microscope (FESEM), cyclic voltammograms (CV), and electrochemical impedance spectra (EIS). Experiment parameters were optimized. Under the optimized condition, the prepared NiS-based biosensor exhibited excellent electrocatalytic oxidation of H2O2 and glucose due to their small size. It provided fast and sensitive strategy for detecting H2O2 and glucose in the range of 1–5000 and 1–1000 μM. The detection limit of 0.257 and 0.3 μM was obtained for H2O2 and glucose. The mechanisms were also analyzed. The proposed biosensor exhibited excellent anti-interference and repeatability. Furthermore, it was applied in the actual sample analysis, such as human blood serum.