The output characteristics on various rectification methods of X-ray equipments for the purpose of medical use are investigated by means of X-ray multimeter. in comparison the measured values with setting values, the accuracy of X-ray tube voltage and...
The output characteristics on various rectification methods of X-ray equipments for the purpose of medical use are investigated by means of X-ray multimeter. in comparison the measured values with setting values, the accuracy of X-ray tube voltage and illumination time was varied in the range of +70% in case of three phase type including inverter type. The inverter type also showed a good reproducibility. As the experimental results, it is desirable that the inverter type X-ray equipment will be widely used in the near future.