    To investigate the minimum near-infrared ray intensity required (quantifiable threshold value) for consistent measurements of axial length (AL) using partial coherence interferometry (PCI), we attached two polarizing lenses (PL) to two types of PCI (IOLmaster, ALscan). The near-infrared ray intensity of PCI was modified by rotating the axis of one PL at intervals of 5 degrees. The right eye of each volunteer was measured three times and the AL and signal-to-noise ratio (SNR) was recorded five times for each measurement. Reduction of light intensity was theoretically estimated using Malus’ Law. AL was measured consistently with both IOLmaster and ALscan until they reached 55 degrees (1.33 % of intensity) and 60 degrees (0.77%), respectively (P = 0.343, Log-rank test). In contrast, SNR decreased as light intensity decreased. In addition, to analyze media opacities that precluded measurement of AL, we retrospectively reviewed the medical records of patients unmeasurable by PCI (ALscan) from May to November 2013.
    Thirty-eight of 473 eyes (8.0%) could not be measured using ALscan due to media opacities, such as severe posterior subcapsular cataract (PSC, 11 eyes), hypermature cataract (9 eyes), and vitreous hemorrhage (18 eyes). The mean grades of vitreous haze and PSC were 7.72 ± 0.96 and 4.45 ± 1.04, respectively. In conclusion, up to 0.77-1.33% of near-infrared rays decreased, and AL could be measured consistently.
    번역하기

    To investigate the minimum near-infrared ray intensity required (quantifiable threshold value) for consistent measurements of axial length (AL) using partial coherence interferometry (PCI), we attached two polarizing lenses (PL) to two types of PCI (I...

    To investigate the minimum near-infrared ray intensity required (quantifiable threshold value) for consistent measurements of axial length (AL) using partial coherence interferometry (PCI), we attached two polarizing lenses (PL) to two types of PCI (IOLmaster, ALscan). The near-infrared ray intensity of PCI was modified by rotating the axis of one PL at intervals of 5 degrees. The right eye of each volunteer was measured three times and the AL and signal-to-noise ratio (SNR) was recorded five times for each measurement. Reduction of light intensity was theoretically estimated using Malus’ Law. AL was measured consistently with both IOLmaster and ALscan until they reached 55 degrees (1.33 % of intensity) and 60 degrees (0.77%), respectively (P = 0.343, Log-rank test). In contrast, SNR decreased as light intensity decreased. In addition, to analyze media opacities that precluded measurement of AL, we retrospectively reviewed the medical records of patients unmeasurable by PCI (ALscan) from May to November 2013.
    Thirty-eight of 473 eyes (8.0%) could not be measured using ALscan due to media opacities, such as severe posterior subcapsular cataract (PSC, 11 eyes), hypermature cataract (9 eyes), and vitreous hemorrhage (18 eyes). The mean grades of vitreous haze and PSC were 7.72 ± 0.96 and 4.45 ± 1.04, respectively. In conclusion, up to 0.77-1.33% of near-infrared rays decreased, and AL could be measured consistently.

    더보기

    참고문헌 (Reference)

    1 L. T. Chylack, Jr., "The lens opacities classification system III. The longitudinal study of cataract study group" 111 : 831-836, 1993

    2 T. Olsen, "Sources of error in intraocular lens power calculation" 18 : 125-129, 1992

    3 T. S. Dietlein, "Signal quality of biometry in silicone oil-filled eyes using partial coherence laser interferometry" 31 : 1006-1010, 2005

    4 J. L. Davis, "Scale for photographic grading of vitreous haze in uveitis" 150 : 637-641e1, 2010

    5 T. Ueda, "Relationship between postoperative refractive outcomes and cataract density: Multiple regression analysis" 36 : 806-809, 2010

    6 W. Drexler, "Partial coherence interferometry:a novel approach to biometry in cataract surgery" 126 : 524-534, 1998

    7 O. Findl, "Optical biometry in cataract surgery" 34 : 131-140, 2002

    8 W. Li, "Novel absolute displacement sensor with wide range based on malus law" 9 : 10411-10422, 2009

    9 R. N. Weinreb, "Measurement of the magnitude and axis of corneal polarization with scanning laser polarimetry" 120 : 901-906, 2002

    10 I. Damian, "Malus’ law for a real polarizer"

    1 L. T. Chylack, Jr., "The lens opacities classification system III. The longitudinal study of cataract study group" 111 : 831-836, 1993

    2 T. Olsen, "Sources of error in intraocular lens power calculation" 18 : 125-129, 1992

    3 T. S. Dietlein, "Signal quality of biometry in silicone oil-filled eyes using partial coherence laser interferometry" 31 : 1006-1010, 2005

    4 J. L. Davis, "Scale for photographic grading of vitreous haze in uveitis" 150 : 637-641e1, 2010

    5 T. Ueda, "Relationship between postoperative refractive outcomes and cataract density: Multiple regression analysis" 36 : 806-809, 2010

    6 W. Drexler, "Partial coherence interferometry:a novel approach to biometry in cataract surgery" 126 : 524-534, 1998

    7 O. Findl, "Optical biometry in cataract surgery" 34 : 131-140, 2002

    8 W. Li, "Novel absolute displacement sensor with wide range based on malus law" 9 : 10411-10422, 2009

    9 R. N. Weinreb, "Measurement of the magnitude and axis of corneal polarization with scanning laser polarimetry" 120 : 901-906, 2002

    10 I. Damian, "Malus’ law for a real polarizer"

    11 R. W. Knighton, "Linear birefringence of the central human cornea" 43 : 82-86, 2002

    12 Y. T. Hsieh, "Intraocular lens power measured by partial coherence interferometry" 89 : 1697-1701, 2012

    13 T. Ueda, "Impact of nuclear cataract density on postoperative refractive outcome: IOL master versus ultrasound" 221 : 384-387, 2007

    14 O. Findl, "High precision biometry of pseudophakic eyes using partial coherence interferometry" 24 : 1087-1093, 1998

    15 Y. A. Chen, "Evaluation of 2new optical biometry devices and comparison with the current gold standard biometer" 37 : 513-517, 2011

    16 K. Irsch, "Birefringence of the central cornea in children assessed with scanning laser polarimetry" 17 : 086001-, 2012

    17 N. Hirnschall, "Assessment of a new averaging algorithm to increase the sensitivity of axial eye length measurement with optical biometry in eyes with dense cataract" 37 : 45-49, 2010

    18 A. R. Wissa, "Agreement and relationship between ultrasonic and partial coherence interferometry measurements of axial length and anterior chamber depth" 6 : 193-198, 2012

    19 P. J. Buckhurst, "A new optical low coherence reflectometry device for ocular biometry in cataract patients" 93 : 949-953, 2009

    20 J. Santodomingo-Rubido, "A new non-contact optical device for ocular biometry" 86 : 458-462, 2002

    더보기

    분석정보

    View

    상세정보조회

    0

    Usage

    원문다운로드

    0

    대출신청

    0

    복사신청

    0

    EDDS신청

    0

    동일 주제 내 활용도 TOP

    더보기

    주제

    연도별 연구동향

    연도별 활용동향

    연관논문

    연구자 네트워크맵

    공동연구자 (7)

    유사연구자 (20) 활용도상위20명

    인용정보 인용지수 설명보기

    학술지 이력

    학술지 이력
    연월일 이력구분 이력상세 등재구분
    2023 평가 해외DB학술지평가 신청대상 (해외등재 학술지 평가)
    2020-01-01 등재 등재학술지 유지 (해외등재 학술지 평가) KCI등재
    2017-02-03 학술지명변경 한글명 : Journal of the Optical Society of Korea -> Current Optics and Photonics
    외국어명 : Journal of the Optical Society of Korea -> Current Optics and Photonics
    KCI등재
    2010-01-01 등재 등재학술지 유지 (등재유지) KCI등재
    2009-01-02 학술지명변경 한글명 : Journal of Optical Society of Korea -> Journal of the Optical Society of Korea
    외국어명 : Journal of Optical Society of Korea -> Journal of the Optical Society of Korea
    KCI등재
    2008-01-01 등재 등재학술지 유지 (등재유지) KCI등재
    2005-01-01 등재 등재학술지 선정 (등재후보2차) KCI등재
    2004-01-01 등재 등재후보 1차 PASS (등재후보1차) KCI등재후보
    2003-01-01 등재 등재후보학술지 선정 (신규평가) KCI등재후보
    더보기

    학술지 인용정보

    학술지 인용정보
    기준연도 WOS-KCI 통합IF(2년) KCIF(2년) KCIF(3년)
    2016 0.67 0.24 0.55
    KCIF(4년) KCIF(5년) 중심성지수(3년) 즉시성지수
    0.48 0.43 0.383 0.02
    더보기

    이 자료와 함께 이용한 RISS 자료

    나만을 위한 추천자료

    해외이동버튼