In this experiment, how DEXA(Dual-energy X-ray Absorptiometry) bone mineral density was measured using the equipment. In order to maintain the same measurement conditions, bone mineral density measurements of 10 cm thick phantom, with an actual patien...
In this experiment, how DEXA(Dual-energy X-ray Absorptiometry) bone mineral density was measured using the equipment. In order to maintain the same measurement conditions, bone mineral density measurements of 10 cm thick phantom, with an actual patient at a point when examining the same conditions(100 kVp, 1 mA) and then out to the five doses of radiation and its average was calculated by dividing measured. X-ray dose rate measured at the Research Institute, Sword of the gamma survey meters calibrated MEDCOM Ltd. (Inspector GM counter tube) was used, calibration factor is 1.15. On a horizontal plane around the patient, depending on the distance was significantly reduced dose rate. In addition, orientation $0^{\circ}$ head end was higher in the direction of the highest dose rate, $0^{\circ}$ $180^{\circ}$ direction from the direction towards the higher dose rate reduced to some extent in the direction of all the $120^{\circ}$ were able to identify.