http://chineseinput.net/에서 pinyin(병음)방식으로 중국어를 변환할 수 있습니다.
변환된 중국어를 복사하여 사용하시면 됩니다.
증예(症例) : 혈액투석 도관 거치시 발생한 유도도선 (guide wire)의 대퇴정맥 내 삽입
최혜은 ( Hye Eun Choi ),류제영 ( Jay Young Rhew ),선인호 ( In O Sun ),이신은 ( Shin Eun Lee ),이영재 ( Young Jae Lee ),최진화 ( Jin Hwa Choi ),이은영 ( Eun Young Lee ),김정화 ( Jung Wha Kim ),이광영 ( Kwang Young Lee ) 전북대학교 의과학연구소 2006 全北醫大論文集 Vol.30 No.1
When a hemodialysis catheter is placed a guide wire is inserted into a central vein first. The dialysis catheter is advanced over the guide wire. After the insertion of the catheter, the guide wire is removed. We experienced a case of accidental insertion of the entire guide wire, which was not removed by mistake soon after the placement of a temporary hemodialysis catheter into the right femoral vein. After 4 months chest X-ray revealed the guide wire in the right femoral through internal jugular veins. It was removed successfully via nonsurgical percutaneous retrieval with a loop snare with internal jugular approach.