이상철,김호경,전인곤,조명혜,조민형,이수열 대한의용생체공학회 2004 의공학회지 Vol.25 No.2
We developed an x-ray cone-beam micro computed tomography (micro-CT) system for small-animal imaging. The micro-CT system consists of a 2-D flat-panel x-ray detector with a field-of-view (FOV) of 120120 mm2, a micro-focus x-ray source, a scan controller and a parallel image reconstruction system. Imaging performances of the micro-CT system have been evaluated in terms of contrast and spatial resolution. The minimum resolvable contrast has been found to be less than 36 CT numbers at the dose of 95 mGy and the spatial resolution about 14 lp/mm. As small animal imaging results, we present high resolution 3-D images of rat organs including a femur, a heart and vessels. We expected that the developed micro-CT system can be greatly used in biomedical studies using small animals.